Agilent 35670A
Step 3. Check the periodic chirp output.
G
• Press the following keys:
[
DC OFFSET
0
[
]
V
[
]
LEVEL
1
[
]
Vpk
[
PERIODIC CHIRP
• Using an an oscilloscope and a 1:1 probe, check the following signal.
Oscilloscope Setup
Connect CH1 to A5 TP8
CH1 V/div
300 V/div
Input Impedance
50 Ω
CH1 Coupling
dc
Probe Atten
1
Display Mode
Repetitive
Averaging
off
Time/div
10 µs/div
Trigger
Trg'd
Trig Src
Sweep
Chan1
• If the signal is incorrect, the A5 Analog assembly is probably faulty.
]
]
Parameters
Amplitude
Time
Duty Cycle
Pulse shape
Time
Relationship
Troubleshooting the Analyzer
To troubleshoot source and calibrator failures
Waveform
Periodic Chirp
4-47