Agilent 35670A
Adjusting the Analyzer

To adjust common mode rejection

To adjust common mode rejection
This procedure optimizes the common mode rejection of the A1/A2 Input assemblies.
The standard two channel analyzer has one A1 Input assembly. The optional four
channel analyzer has two A2 Input assemblies: channel 1 and 3 are routed to the A2
Input assembly in the lower slot and channel 2 and 4 are routed to the A2 Input
assembly in the upper slot.
Equipment Required: BNC-to-BNC Cable
BNC(f)-to-Minigrabber Adapter
5-13