Agilent 35670A
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Intermodulation Distortion
Measurement
One Ch, F1+F2, 102.4 kHz, Ch 1
One Ch, F1+F2, 64.096 kHz, Ch 1
One Ch, F1-2F2, 99.096 kHz, Ch 1
Two Ch, F1+F2, 1952 Hz, Ch 1
Two Ch, F1+F2, 1952 Hz, Ch 2
Two Ch, F1-2F2, 1048 Hz, Ch 1
Two Ch, F1-2F2, 1048 Hz, Ch 2
Two Ch, F1+F2, 48.048 kHz, Ch 1
Two Ch, F1+F2, 48.048 kHz, Ch 2
Two Ch, F1+F2, 33.024 kHz, Ch 1
Two Ch, F1+F2, 33.024 kHz, Ch 2
Two Ch, F1-2F2, 49.096 kHz, Ch 1
Two Ch, F1-2F2, 49.096 kHz, Ch 2
Cross Talk
Measurement
Source-to-Ch 1
Source-to-Ch 2
Receiver Ch 1, Driver Ch 2
Receiver Ch 2, Driver Ch 1
Performance Test Record - Two Channel
Lower Limit
Upper Limit
(dBfs)
−80
−80
−80
−80
−80
−80
−80
−80
−80
−80
−80
−80
−80
Lower Limit
Upper Limit
(dBVrms)
−126
−126
−126
−126
11 of 14
Verifying Specifications
Measured Value
Pass/Fail
(dBfs)
Measured Value
Pass/Fail
(dBVrms)