Agilent Technologies 35670A Service Manual page 222

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Troubleshooting the Analyzer
To troubleshoot source and calibrator failures
Step 2. Check the dc offset.
G
• Connect the voltmeter to A5 TP3.
• Press the following keys:
[
]
LEVEL
0
[
]
Vpk
[
DC OFFSET
• Rotate the RPG knob while monitoring the voltmeter.
• If the voltmeter's voltage does not change from +2.3 to 2.3 as the dc offset
value is varied between 10 and +10 Vdc, the A5 Analog assembly is
probably faulty.
4-46
]
Agilent 35670A