Note
Internal Test Descriptions
Calibration Routine Description
[ Inst Mode ]
[ 2 CHANNEL ]
[ Input ]
[ CHANNEL 1 RANGE ]
1
[ dBVrms ]
[ CHANNEL 2 RANGE ]
1
[ dBVrms ]
[ Disp Format ]
[ UPPER/LOWER ]
[ System Utility ]
[ CALBRATIN ]
[ SAVE CH1 CAL TRACE ]
[ INTO D1 ]
[ SAVE CH2 CAL TRACE ]
[ INTO D2 ]
[ Meas Data ]
[ MORE ]
[ DATA REGISTER ]
[ D2 ]
[ Active Trace ]
[ D1 ]
[ Scale ]
[ Y PER DIV (DECADES) ]
1
[ ENTER ]
[ CENTER REFERENCE ]
0
[ ENTER ]
[ Active Trace ]
[ Y PER DIV (DECADES) ]
1
[ ENTER ]
[ CENTER REFERENCE ]
0
[ ENTER ]
Display A shows the message
D2 CAL CHAN 2
.
The calibration correction curves should be contained within
applies to all ranges at full span.
10-8
D1 CAL CHAN 1
and display B shows the message
Agilent 35670A
5 dB. The ±5 dB limit
±