Agilent Technologies 35670A Service Manual page 385

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
Power-on Test Messages
The ''Power-on Test Messages'' table provides additional information for interpreting
the power-on test LEDs. Using the ''Binary to Hexadecimal'' table, translate the
power-on test LEDs to their equivalent hexadecimal code (see ''To troubleshoot
power-up failures'' on page 4-15 for details on decoding the power-on test LEDs to
their binary code). The ''Power-on Test Messages'' table describes the power-on
subtests in the order they are run. The table also shows the relationship between a
failing power-on subtest and the assemblies or sub-blocks.
False error codes can be caused by shorts on the buses, reset line, or interrupt line. If
an error code is caused by the last bus connected, it is probably the source of the
failure.
Binary to Hexadecimal
Binary
1 = LED on
0 = LED off
0000
0001
0010
0011
0100
0101
0110
0111
1000
1001
1010
1011
1100
1101
1110
1111
Hexadecimal
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
Internal Test Descriptions
Power-on Test Description
10-3