Verifying Specifications
Operation Verification Test Record - Two Channel
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Measurement
Ch 1, 100 Hz
Ch 2, 100 Hz
Ch 1, 1000 Hz
Ch 2, 1000 Hz
Ch 1, 10000 Hz
Ch 2, 10000 Hz
Ch 1, 25120 Hz
Ch 2, 25120 Hz
Channel Match
Measurement
Two Ch, 2/1, 7 dBV FS Mag
Two Ch, 2/1, 7 dBV FS Phs
Two Ch, 2/1, −13 dBV FS Mag
Two Ch, 2/1, −13 dBV FS Phs
Two Ch, 2/1, −33 dBV FS Mag
Two Ch, 2/1, −33 dBV FS Phs
Two Ch, 2/1, 7 dBV −20dBfs Mag
Two Ch, 2/1, 7 dBV −20dBfs Phs
Lower Limit
Upper Limit
(dB)
(dB)
0.7
−0.7
0.7
−0.7
0.7
−0.7
0.7
−0.7
2
−3
2
−3
2.4
−4.5
2.4
−4.5
Lower Limit
Upper Limit
0.04 dB
−0.04 dB
0.5 deg
−0.5 deg
0.04 dB
−0.04 dB
0.5 deg
−0.5 deg
0.04 dB
−0.04 dB
0.5 deg
−0.5 deg
0.08 dB
−0.08 dB
0.5 deg
−0.5 deg
8 of 10
Agilent 35670A
Measured Value
Pass/Fail
(dB)
Measured Value
Pass/Fail
dB
deg
dB
deg
dB
deg
dB
deg