Turning the Power ON ( | ) and OFF ( O )
<Turning the Power ON>
• The measurement screen appears after the
start-up screen.
[Operating Procedure]
Set the POWER switch to ( | ).
1
<Turning the Power OFF>
[Operating Procedure]
1
Set the POWER switch to ( O ).
• After turning on the power to the measuring head, turn on the power to the data processor.
If a communication error occurs, press the Escape key.
• To turn OFF the power switch, observe the following instructions:
1) Do not turn OFF the data processor power switch immediately after execution of mea-
surement (including calibration and target color measurement for update).
2) Do not turn OFF the data processor power switch when transferring data from the mea-
suring head.
3) Do not turn OFF the measuring head power switch when automatic averaging measure-
ment is in progress with the data processor.
In the above cases 1) and 2), the data processor receives measurement data from the measur-
ing head, and stores the data into the data processor internal memory after necessary process-
ing. If the power switch is turned OFF before completion of data storage, the data may be lost,
or incompletely stored. If an incomplete storing condition is detected at the next startup, the
data processor may initialize all stored data because it recognizes corruption of stored data.
(For initialization, see NOTE in <Data Memory and Memory Backup> on P.66.) In case of
the above 3), the data processor may not accept key entry even after the measuring head is re-
started and re-connected. In such a condition, restart the data processor.
Power switch
Power switch
65