Konica Minolta CR-400 Instruction Manual page 85

Chroma meter
Hide thumbs Also See for CR-400:
Table of Contents

Advertisement

Color difference graph
The display is as shown in the figure at right when no color difference
tolerance is set.
Setting the color difference tolerance (see P.116 "Setting the Color
Difference Tolerance") makes the following color difference graph
display is displayed.
1) ∆L* axis
2) ∆a* axis
3) ∆b* axis
4) Measurement point
5) Color difference target color measurement point
6) Displaying a box-type color difference tolerance
7) Scale on the ∆L* and ∆b* axes (Indicates a value at the minus end point of each axis. The scale
setting will be automatically changed.)
8) Displaying a elliptical color difference tolerance
O PASS: This is displayed when the value is within
the color difference tolerance.
∆ WARN: This is displayed when value is within the
color difference tolerance, but exceeds the
warning level.
X FAIL: This is displayed when the value exceeds
the color difference tolerance.
Box-type color difference
tolerance setting
Within the color
difference tolerance
Out of the color
difference tolerance
[Setting conditions]
• The color difference target color
is set.
• L*a*b* is selected by pressing the
Color Space key
• The color difference graph is se-
lected by pressing the Change
Display key
.
Elliptical color difference
tolerance setting
Within the color
difference tolerance
Out of the color
difference tolerance
.
83