Konica Minolta CR-400 Instruction Manual page 136

Chroma meter
Hide thumbs Also See for CR-400:
Table of Contents

Advertisement

User Calibration
Input range: 0.01 ≤ X, Y, Z ≤ 160
8
Move the cursor using the
using the numeric pad.
9
Place the measuring head vertically above a target
specimen.
• Do not move the measuring head during measurement.
• Through user calibration for an existing calibration channel, the user calibration data will
be overwritten when the measurement is completed.
10
Press the Measure/Enter key
head) after making sure the ready lamp is ON.
• Skip to step 12 if not setting a calibration channel name.
Once measurement is complete, select the "CAL.Ch NAME"
11
using the
the numeric pad.
(How to set the calibration channel name)
1) Select input mode using the
• The input modes change in the following order: upper case Latin letters → lower case
Latin letters → numbers.
2) Move the cursor using the
the numeric pad.
• The name is up to 9 characters (see P.20 "As input keys" for details.)
• Settings are complete, complete sound beeps and the display returns to the measurement screen.
12
Once settings are complete, press the Measure/Enter key
134
key and set the value
key and set the name using the
key.
key and input the color difference target color name using
(or measurement button on the measuring
key and
Setting screen
Setting screen
.