Elliptical tolerance

This judges whether or not the value of the measured color difference is within the range of the ellip-
tical tolerance set for the color difference target color. When setting an elliptical tolerance, besides the
tolerance value, an offset and a rotation can also be set.
If the set color space is "L*C*h" the col or difference tolerance is set along the "L*a*b*" color space
axis, and if the set color space is "L99C99h99" then the "L99a99b99" color space axis is used.
See P.80 "Color Space and Color Difference Setting" for details.
Set a color difference target color ahead of time before setting an elliptical limit.
Since the color difference graphs are displayed in two dimensions, the fail judgment data may some-
times be displayed as plotted inside the tolerance in each graph.
Selecting the limit value in P.115 displays the [limit value] selection screen.
5,7
[Operating Procedure]
Use the
key to select a color difference target
1
color.
• The tolerance type screen appears.
2
Press the Measure/Enter key
• The limit value/offset setting screen appears.
Select the "ELLIPTICAL" as tolerance type.
3
• If a color difference tolerance is already set, go to step 6 as the
set color difference tolerance data is displayed.
Press the Measure/Enter key
4
.
.
1,3,5,7
2,4,6,8
Selection screen
Tolerance type screen
Setting screen
117