Konica Minolta CR-400 Instruction Manual page 142

Chroma meter
Hide thumbs Also See for CR-400:
Table of Contents

Advertisement

User Calibration
6
Select either one channel selected or "ALL CHAN-
NELS" using the
• The deletion screen for calibration appears.
Press the Measure/Enter key
7
• Once deletion is complete, complete sound beeps and the display returns to the measurement
screen.
White calibration is selected if the selected calibration channel is deleted.
Confirm the message and press the Measure/Enter key
8
(Method for changing user calibration channel name)
1) Use Steps 1-7 of the operating procedure "Setting the User Calibration
Data" on P.132 to display the setting screen for the calibration channel to
be changed.
• In Step 4, from the existing calibration channels, select the one you want
to change the name of.
2) With the
3) With the
• The input mode changes in the sequence: Upper case letters → Lower case letters
4) Move the cursor with the
with the number keys.
• The name may be up to 9 characters long. (For details, see "As input keys " on P.20.)
5) When the setting is complete, press the Measure/Enter key
• If the calibration channel name has been changed, measurement is not made even if the
Measure/Enter key
→Numbers.
• After the settings are complete, the completion sound is made and the system returns to
the measuring screen.
140
key.
key, select "CAL. Ch NAME".
key, select the input mode.
key and input the user calibration channel name you want
is pressed.
.
Only deleting the
selected channel
.
.