Konica Minolta CR-400 Instruction Manual page 89

Chroma meter
Hide thumbs Also See for CR-400:
Table of Contents

Advertisement

After the initial setting is complete, the illuminant is set to C, so set the data for C. When using
D
, change the setting to "D
Setting 5) Illuminant". After this, return to this page and set the data for D
• See P.76 "Initial Setting" and P.72 if illuminant is changed after calibration.
65
The white calibration plate is placed near the center.
Use the center area when white calibrating.
3
Place the measuring head vertically above the middle of the white calibration
plate.
• Calibration is complete after the lamp flashes three times. Complete sound beeps and the
display returns to the measurement screen.
4
Press the Measure/Enter key
head) after making sure the ready lamp is ON.
The illuminant lamp
flashes three times.
• Do not move the measuring head during calibration.
• If selected in user calibration, select "white calibration plate" following steps 1 to 4 on
P.137 "Setting the Manual Select Calibration" to set to the white calibration.
With this, white calibration is finished.
When measuring the chroma values (absolute values), perform P.88 "Absolute Measurement". To
measure the color difference between a color difference target color and a specimen, perform P.90
"Setting the Color Difference Target Color" and P.98 "Measuring the Color Difference".
" before calibration as described in steps 1 to 4 on P.72 "Basic
65
(or measurement button on the measuring
.
65
Measurement screen
87