Set-Up Procedures (Cont.)
Digital Display Module-
Two additional levels of "advanced" set-up
features are available.
The first advanced level, S S E E T T 2 2 , is used to
adjust communication related parameters.
CAUTION:
Communication parameters are factory set and
should not be changed. These instructions are
provided for programming replacement modules.
Changing Communication Parameters
A A D D D D R R (network address)
A network address uniquely identifies each
digital display module in the alarm panel.
Address range is 0 0 to 2 2 5 5 5 5 .
For TotalAlert 2 alarm panels, this setting is
programmed at the factory when the alarm
panel is assembled and should not be
changed.
B B A A U U D D (baud rate)
Used to set communication speed. User
choices are 4 4 8 8 0 0 0 0 or 9 9 6 6 0 0 0 0 .
Baud rate must be set to 9 9 6 6 0 0 0 0 . for
TotalAlert 2 alarm panels.
N N D D F F (network data format)
Defines data word length. Choices are 8 8
(bits) or 1 1 6 6 (bits). NDF must be set to 1 1 6 6
for TotalAlert 2 alarm panels.
P P D D F F (pressure data format)
Defines data word length transmitted to a
slave alarm panel. User choices are 8 8
(bits) or 1 1 6 6 (bits). Factory setting is 1 1 6 6 .
66
6-847718-00 Rev. B00
1. Press and HOLD SETUP button for
2. Press and release TEST button on
Advanced
3. Press and release TEST button.
4. Press and release TEST button. N N D D F F
5. Press and release TEST button. P P D D F F
6. Press and release TEST button to
7. Digital display module will return to
TWO seconds until S S E E T T 2 2 is displayed.
front of alarm panel. A A D D D D R R will be
displayed for one second and then
current network address. Select
desired address by pressing
buttons.
Each module must have a unique
address.
B B A A U U D D will be displayed for one second
and then current baud rate. Select
desired baud rate by pressing
buttons.
will be displayed for one second and
then current network data format.
Select desired data format by
pressing
or
buttons.
will be displayed for one second and
then current pressure data format.
Select desired data format by
pressing
or
buttons.
save changes. D D O O N N E E will be
displayed for two seconds.
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will
default to previous settings. No changes to
settings will be saved.
Set-up must be completed and module must
D D O O N N E E
display
in order for changes to be
saved.