Set-Up Procedures (Cont.)
Digital Display Module-
(Cont.)
The second advanced level, S S E E T T 3 3 , is
used to adjust other miscellaneous
parameters as listed below:
Changing Miscellaneous Parameters
S S E E R R U U (gas service identification)
Used to set digital display module's gas
service identification for communication
with a digital sensor.
User choices are:
0 0 2 2
(oxygen)
N N 2 2 0 0
(nitrous oxide)
A A , , R R
(medical air)
U U A A ( (
(vacuum)
N N 2 2
(nitrogen)
L L A A , , R R
(lab air)
L L U U A A ( (
(lab vacuum)
A A R R 9 9 N N
(argon)
I I A A , , R R
(instrument air)
U U A A 9 9 D D
(WAGD)
( ( 0 0 2 2
(carbon dioxide)
( ( - - 0 0 2 2
(CO2/O2)
0 0 2 2 - - ( (
(O2/CO2)
K K E E 0 0 2 2
(He/O2)
0 0 2 2 K K E E
(O2/He)
K K E E
(Helium)
N N O O N N E E
(none)
O O F F F F S S (display offset)
Used to adjust digital module's display.
User choices are 0 0 , 1 1 , 2 2 , - - 1 1 , or - - 2 2 .
Factory setting is 0 0 .
R R E E P P T T (audible alarm repeat)
Used to set time interval for reactivation of
audible alarm after alarm MUTE
has been pressed.
User choices are:
O O F F F F
1 1 , 5 5 , 1 1 5 5 , 3 3 0 0 (minutes), 1 1 H H R R , 8 8 H H R R , 1 1 2 2 H H R R or
Advanced
2 2 4 4 H H R R (hour).
Factory setting is O O F F F F .
When set to O O F F F F , audible alarm will not
reactivate after an active alarm has been
silenced. Any new alarm will reactivate
audible alarm.
I I N N T T S S (display intensity)
Used to adjust digital module's display
brightness.
Factory setting is maximum intensity.
1. Press and HOLD SETUP button for
2. Press and release TEST button on front
3. Press and release TEST button. O O F F F F S S
4. Press and release TEST button. R R E E P P T T
5. Press and release TEST button. I I N N T T S S
6. Press and release TEST button to save
button
FOUR seconds until S S E E T T 3 3 is displayed.
of alarm panel. S S E E R R U U will be displayed
for one second and then current gas
service identification. Select desired
gas service by pressing
buttons.
will be displayed for one second and
then current display offset. Select
offset by pressing
will be displayed for one second and
then current audible alarm repeat time.
Select desired repeat time by pressing
or
buttons.
will be displayed. Select desired
display brightness by pressing
buttons. Each push will change
brightness of display. Audible alarm will
chirp one time with each push and two
times when limit reached.
changes. D D O O N N E E will be displayed for
two seconds.
or
or
buttons.
or