Wiring Schematic 4: Relay Board

Relay
board A
Breakout
board A
NOTE:
Field wiring cable shields must be grounded at
only one end, inside alarm panel back box. Refer
to page 39 for details.
Relay
board B
Breakout
board B
SIGNAL A1
SIGNAL A2
SIGNAL A32
SIGNAL B1
SIGNAL B32
Dry normally-closed
output relay contacts
NOTE:
Output relay contacts are dry normal-
ly closed type. Contacts open upon
designated alarm activation.
Pressing MUTE
on Annunciator
Module DOES NOT reset relay.
Relay deactivates when alarm condi-
tion is corrected.
Relay contact ratings are 2 A @ 30
VDC/0.5 A @ 125 VAC.
6-847718-00 Rev. B00
49