Wiring (Cont.)

Field Wiring Cable Shield Grounding

All field wiring cable shields must be
grounded inside alarm panel back box.
A grounding screw is provided in upper
right corner of area alarm back box
(Figure 28).
Grounding screws are provided in top and
bottom of master and combo alarm back
boxes (Figure 29).
Shields from several cables may be
wrapped together and crimped into one
ring lug (by others) .
CAUTION:
Keep shield wires to ground screw as short as
possible so they can not touch front panel circuit
boards when front panel is closed.
Cable shield
grounding screw
(crimp lug not provided)
Figure 28: Cable Shield Grounding-Area
grounding screws
located at top and
bottom of back box
(crimp lugs not provided)
Figure 29: Cable Shield Grounding-Master/Combo
Cable shield
6-847718-00 Rev. B00
39