Pull Wire
General Requirements
•
Power all alarms from life safety
branch of emergency power system as
required by applicable standards.
•
Protect all wiring from physical
damage by raceways or conduit as
required by applicable standards.
•
Wire master panels directly to
switches or sensors as required by
applicable standards.
•
Wiring runs should be made with color
coded wire. Record color, signal, and
source of signal for each wiring lead to
aid in connection of alarm finish
components.
•
Avoid installing alarm panels near
radio transmitters, electrical motors, or
switchgear.
Wire Type And Size
All low voltage wiring must meet the fol-
lowing criteria:
•
Copper wire no smaller than 22 AWG.
•
Circuit length not to exceed 5000 feet.
•
Cable must be twisted-pair shielded
type. Multi-pair cables within one
common shield are acceptable.