Pull Wire; General Requirements; Wire Type And Size - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Pull Wire

General Requirements

Power all alarms from life safety
branch of emergency power system as
required by applicable standards.
Protect all wiring from physical
damage by raceways or conduit as
required by applicable standards.
Wire master panels directly to
switches or sensors as required by
applicable standards.
Wiring runs should be made with color
coded wire. Record color, signal, and
source of signal for each wiring lead to
aid in connection of alarm finish
components.
Avoid installing alarm panels near
radio transmitters, electrical motors, or
switchgear.

Wire Type And Size

All low voltage wiring must meet the fol-
lowing criteria:
Copper wire no smaller than 22 AWG.
Circuit length not to exceed 5000 feet.
Cable must be twisted-pair shielded
type. Multi-pair cables within one
common shield are acceptable.
27
6-847718-00 Rev. B00