Combo Alarms - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Alarm Configurations (Cont.)

Combo Alarms

Combo alarms include the following modular
components:
Annunciator module with VFD
One multiplexer module
One LED module (Large Combo Only).
One or two breakout boards
One or two relay boards (Optional)
3 or 6 digital display or blank modules
Combo alarms can monitor 32 or 64
switched inputs.
Inputs can be assigned to any one of 5
(small combo) or 10 (large combo) gas
service indicators.
Optional dry contact relays are available
for all signals.
Combo alarms can also monitor up to 3
(small combo) or 6 (large combo) digital
display modules.
NOTE:
Sensors on combo alarms are always remotely
mounted.
Model Number Scheme:
6-TA2C _ _ _ - _ _ _ _ _ _
32 or 64
Designates number
of signals
14
6-847718-00 Rev. B00
Position within alarm panel
(Figures 11 and 12)
1 2 3 4 5 6
Designates type of alarm
panel component from table
R or blank
R = With Relays
O
OXYGEN
D
OXYGEN - 100 PSI
X
NITROUS OXIDE
A
MEDICAL AIR
F
MEDICAL AIR - 100 PSI
5
AIR (ISO)
7
LABORATORY AIR
2
O2-CO2
4
O2-He
C
CARBON DIOXIDE
G
CARBON DIOXIDE - 100 PSI
1
CO2-O2
3
He-O2
H
HELIUM
V
VACUUM
6
VACUUM (ISO)
8
LABORATORY VAC
W
WAGD
N
NITROGEN
9
INSTRUMENT AIR
J
ARGON
B
BLANK
NOTE:
Multi-signal modules are not allowed in
combination alarms.
1 2 3
Figure 11: Small Combo Alarm Front Panel
1 2 3
4 5 6
Figure 12: Large Combo Alarm Front Panel

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents