Start-Up and Checking (Cont.)
Combo Alarm Panel
1. Turn on electrical power to alarm
panel.
2. Make following observations:
•
POWER ON light illuminates on
front of alarm panel.
•
Boot message is shown on
vacuum fluorescent display.
Booting...
BeaconMedaes
TotalAlert 2 System
Version: [VERSION]
SN 12345
Browse at:
TA2_12345
•
Alarm modules warm-up for ten
seconds. During warm-up, no
audible or visual alarms activated.
After warm-up, audible and visual
alarms are initiated for any active
alarm.
3. Press MUTE
audible alarm.
4. Pressurize piping system (medical gas
and vacuum).
5. Make following observations:
•
Digital display modules display
actual pipeline pressures or
vacuum levels.
•
If digital display module pipeline
pressures and vacuum levels are
within pre-set limits, NORMAL
indicator is illuminated.
•
If digital display module pipeline
pressures and vacuum levels are
outside pre-set limits, status
indicators show HIGH or LOW.
If a fault code is displayed (- - F F 1 1 - - ,
•
- - F F 2 2 - - , - - F F 3 3 - - or - - F F 4 4 - - ) refer to
Troubleshooting Guide.