Combo Alarm Panel - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Start-Up and Checking (Cont.)

Combo Alarm Panel

1. Turn on electrical power to alarm
panel.
2. Make following observations:
POWER ON light illuminates on
front of alarm panel.
Boot message is shown on
vacuum fluorescent display.
Booting...
BeaconMedaes
TotalAlert 2 System
Version: [VERSION]
SN 12345
Browse at:
TA2_12345
Alarm modules warm-up for ten
seconds. During warm-up, no
audible or visual alarms activated.
After warm-up, audible and visual
alarms are initiated for any active
alarm.
3. Press MUTE
audible alarm.
4. Pressurize piping system (medical gas
and vacuum).
5. Make following observations:
Digital display modules display
actual pipeline pressures or
vacuum levels.
If digital display module pipeline
pressures and vacuum levels are
within pre-set limits, NORMAL
indicator is illuminated.
If digital display module pipeline
pressures and vacuum levels are
outside pre-set limits, status
indicators show HIGH or LOW.
If a fault code is displayed (- - F F 1 1 - - ,
- - F F 2 2 - - , - - F F 3 3 - - or - - F F 4 4 - - ) refer to
Troubleshooting Guide.
58
6-847718-00 Rev. B00
http://
button to silence
Multiplexer modules indicate current
status from dry-contact switches.
Connected, but unprogrammed
signals are shown on the vacuum
fluorescent display.
Unassigned
Input [INPUT NUMBER]