Set-Up Procedures; Annunciator Module - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Set-Up Procedures

Annunciator Module

Setup Using Front Panel Controls
(Master and combo alarm panels)
Many parameters may be set using front
panel controls, but some parameters must
be set up using web pages.
NOTE:
Before setting up signals, fill out Master Alarm
Signal Input Data table on page 138.
1. Press SETUP button on back of
annunciator module to enter setup
mode.
****
Device
****
Setup
Press 'TEST' for
the next item
Press
or
to
change the value
2. N2 Address
N2 Address
[VALUE]
This parameter is the Johnson Controls
Metasys® N2 Address.
Possible values are:
1- 255 (Factory default: 1)
3. Clear Network
Clear network
[YES/NO]
Answering 'Yes' will clear and rebuild
the medical gas network.
4. Reset Users
Reset users
[YES/NO]
Answering Yes will clear all user ids
and reset User 1 name and password
to factory defaults:
Username: new
Password: new
****
****
5. Input gas type
Input XXX gas type
[GAS TYPE]
Sets the gas type for signal XXX.
Possible GAS TYPES are:
Oxygen
Nitrous Oxide
Medical Air
Vacuum
Nitrogen
Lab Air
Lab Vacuum
Argon
Instrument Air
WAGD
Carbon Dioxide
CO2-O2
O2-CO2
He-O2
O2-He
Helium
Custom
Unused
Dental Air
If gas type is set to 'Unused', step 5 is
repeated for the next input number.
6. Input Message
Input XXX message
[Message]
Sets the alarm message for signal
XXX.
There are several pre-programmed
messages or a custom message can
be selected. A custom message must
be entered using web page setup.
Possible values depend on gas type
selected in previous step (See Table 1).
7. Input lights LED
Input XXX lights
LED [LED Number]
Assigns LED to signal XXX.
Possible values are 1 - 10 for master
and large combo alarms, and 1 - 5 for
small combo alarms.
59
6-847718-00 Rev. B00