Set-Up Procedures (Cont.)
Multiplexer Module (Cont.)
CAUTION:
Communication parameters are factory set and
should not be changed. These instructions are
provided for programming replacement modules.
Setting Communication Parameters:
Address range is 0 to 255. Factory default
setting is 8.
Network address is shown as a binary
number on specific red and green front
panel LEDs. A numeric value is assigned
to each LED (Figure 45).
For example, if LEDs for bits 0 and 2 were
illuminated, network address would be 5 (1
+ 4 = 5).
1. Press and HOLD set-up button.
Audible alarm will chirp when set-up
button is first pushed and again two
seconds later. Release button after
second audible alarm chirp.
2. Multiplexer module front panel LEDs
will show network address. If no LEDs
are illuminated, address is 0.
3. Using either
or
desired network address between 0
and 255.
4. Press and release TEST button to save
changes. LEDs will sweep from top to
bottom to indicate end of set-up
sequence.
5. Multiplexer module will return to
monitoring mode.
buttons, select
Red LEDs
LED 1
Bit 0 (1)
LED 2
Bit 1 (2)
LED 3
Bit 2 (4)
LED 4
Bit 3 (8)
LED 5
Figure 45: Binary data represented by LEDs
Green LEDs
Bit 4 (16)
Bit 5 (32)
Bit 6 (64)
Bit 7 (128)
63