Multi-Signal Module

Press and hold front panel TEST button for
5 s to initiate following sequence:
Audible alarm sounds for 1 s.
All red and green LEDs flash for 5 s.
Software revision is displayed as
follows:
-
Expressed as a three digit number
(e.g. 1.0.2).
-
Each digit of number is displayed
separately for 2 s and separated
from the next digit by a brief flashing
of all indicators.
-
Each number is displayed in a
binary code where each multi-signal
module LED has a binary value
(Figure 110.).
-
Example: Software revision 1.0.2
would be displayed as follows:
First digit is determined by 2 s
illumination of top left indicator
(bit 0 = number 1).
All LEDs quickly flash.
Second digit is determined to be
0 because no indicators are
illuminated for 2 s.
All LEDs quickly flash.
Third digit of is determined by 2
s illumination of 2nd red indicator
(Bit 1 = number 2).
After all three digits of software revision
have been displayed, all LEDs
illuminate for 1 s before displaying the
module's network address.
Network address is displayed for 2 s in
a binary code where each multi-signal
module LED has a binary value
(Figure 110.).
118
6-847718-00 Rev. B00
All LEDs illuminate for 1 s and module
returns to monitoring mode.
Red LEDs
SIGNAL 1
Bit 0 (1)
SIGNAL 2
Bit 1 (2)
SIGNAL 3
Bit 2 (4)
SIGNAL 4
Bit 3 (8)
SIGNAL 5
Figure 110: Binary data represented by LEDs
Green LEDs
Bit 4 (16)
Bit 5 (32)
Bit 6 (64)
Bit 7 (128)