BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions page 19

Table of Contents

Advertisement

Electromagnetic Compatibility (Cont.)
EN 60601-1-2 (Cont.)
Guidance and manufacturer's declaration - electromagnetic emissions
The TotalAlert 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the TotalAlert 2
should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The TotalAlert 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the TotalAlert 2
should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
±6 kV contact
Discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
U
Voltage dips, short
<5 %
T
Interruptions and
(>95 % dip in
voltage variations
for 0,5 cycle
on power supply
U
<40 %
input lines
T
(>60 % dip in
IEC 61000-4-11
for 5 cycles
U
<70 %
T
(>30 % dip in
for 25 cycles
U
<5 %
T
(>95 % dip in
for 5 sec
Power frequency
3 A/m
(50/60 Hz) magnet-
ic field
IEC 61000-4-8
U
NOTE
is the a.c. mains voltage prior to application of the test level.
T
The TotalAlert 2 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
Group 1
equipment.
The TotalAlert 2 is suitable for use in all establishments other than domestic and those
Class A
directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class A
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
U
<5 %
T
U
U
)
(>95 % dip in
T
for 0,5 cycle
U
<40 %
T
U
U
)
(>60 % dip in
T
for 5 cycles
U
<70 %
T
U
U
)
(>30 % dip in
T
for 25 cycles
U
<5 %
T
U
U
)
(>95 % dip in
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
)
T
TotalAlert 2 requires continued operation during power
mains interruptions, it is recommended that the
TotalAlert 2 be powered from an uninterruptible power
)
supply or battery.
T
)
T
)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
17
6-847718-00 Rev. B00