Set-Up Procedures (Cont.)
Digital Display Module (Cont.)
0 0 to 1 1 3 3 8 8 0 0 (1 1 7 7 2 2 5 5
) (kPa)
*
0 0 to 2 2 0 0 0 0 (2 2 5 5 0 0
)
*
0 0 to 3 3 2 2
0 0 to 8 8 0 0 0 0
*N2 and instrument air only
6. Press and release TEST button.
HIGH indicator will illuminate and K K , ,
will be displayed. After two seconds
current high alarm setting will be
displayed. Select desired high alarm
setting by pressing either
buttons. High alarm setting ranges
are:
0 0 to 1 1 3 3 8 8 0 0 (1 1 7 7 2 2 5 5
) (kPa)
*
0 0 to 2 2 0 0 0 0 (2 2 5 5 0 0
)
*
0 0 to 3 3 2 2 or N N O O N N E E
0 0 to 8 8 0 0 0 0 or N N O O N N E E (mm Hg)
*N2 and instrument air only
7. Press and release TEST button to
save changes. D D O O N N E E will be
displayed for two seconds.
8. Digital display module will return to
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will
default to previous settings. No changes to
settings will be saved.
The set-up must be completed and the module
D D O O N N E E
must display
in order for changes to be
saved.
(psig)
(in Hg)
(mm Hg)
or
(psig)
(in Hg)
65
6-847718-00 Rev. B00