Set-Up Procedures (Cont.)

Digital Display Module

Digital display modules are shipped from
the factory with the following settings:
Gas/Vacuum
Service
Pressure (50 psig)
Pressure (100 psig) 80 psig
Nitrogen, Instr. Air
Vacuum/WAGD
Changing units and alarm set points:
1. Loosen two screws on front of alarm
panel. Hinge open alarm panel's front
cover.
2. Locate SETUP button on circuit board
of digital display module to be adjust-
ed (Figure 46).
3. Press and release SETUP button.
Audible alarm will chirp and S S E E T T 1 1 will
be displayed.
4. Press and release TEST button on
front of alarm panel. Current unit of
measure will be displayed. Select
desired unit of measure by pressing
either
or
of measure are:
K K P P A A
(kPa)
P P S S I I
(psig)
I I N N K K 9 9
(in Hg)
N N N N K K 9 9
(mm Hg)
5. Press and release TEST button. The
LOW indicator will illuminate and L L O O
will be displayed. After two seconds
current low alarm setting will be
displayed. Select desired low alarm
setting by pressing either
buttons. Low alarm setting ranges
are:
64
6-847718-00 Rev. B00
Alarm Set Point
Low
High
40 psig
60 psig
120 psig
140 psig 190 psig
12 in Hg. none
buttons. Available units
or
Digital Display Module Board
SETUP
button
TEST
button
Figure 46: Digital Display Programming Buttons
Low alarm
LED
Display
Figure 47: Digital Display Module
NOTE:
Audible alarm will chirp with each button push.
Up and down
buttons
High alarm
LED