Wiring Schematic 9: Digital Display Module High / Low Relays; Wiring Schematic 8: General Fault / Auxiliary Relays - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Wiring Schematic 8: General Fault / Auxiliary Relays

Auxiliary Relay
N/O (pin 6)
Common (pin 7)
N/C (pin 8)
Annunciator Module

Wiring Schematic 9: Digital Display Module High / Low Relays

High Alarm Relay
Low Alarm Relay
Common (pin 9)
N/C (pin 8)
N/C (pin 10)
Common (pin 7)
Digital Display Module
General Fault Relay
N/C (pin 5)
Common (pin 4)
N/O (pin 3)
20 wire
ribbon cable
to adjacent
module
20 wire
ribbon cable
to adjacent
module
NOTE:
General Fault Relay activates when ANY alarm
on panel is in progress.
Pressing MUTE
on Annunciator Module DOES
NOT reset relay. Relay deactivates when alarm
condition is corrected.
Auxiliary Relay (optional) activates when ANY
audible alarm on panel is in progress.
Pressing MUTE
on Annunciator Module WILL
deactivate relay until audible alarm is again reacti-
vated.
Relay contact ratings are 2A @ 30 VDC/0.5 A @
125 VAC.
NOTE:
High Alarm Relay activates when high line pres-
sure alarm on digital display module is in
progress. When relay activates normally closed
contacts open.
Pressing MUTE
on Annunciator Module DOES
NOT deactivate relay. Relay deactivates when
high line pressure alarm condition is corrected.
Low Alarm Relay activates when low line pressure
alarm on digital display module is in progress.
When relay activates normally closed contacts
open.
Pressing MUTE
on Annunciator Module DOES
NOT deactivate relay. Relay deactivates when
low line pressure alarm condition is corrected.
Relay contact ratings are 2A @ 30 VDC/0.5 A @
125 VAC.
53
6-847718-00 Rev. B00