Alarm System
Periodic testing of alarm system is
recommended. Alarm modules have built-
in self-test modes.
Front panel TEST button will initiate self-
test mode on all alarm panel modules at
same time.
Front panel
TEST button
Figure 108: Alarm Panel Test button
114
6-847718-00 Rev. B00
Annunciator Module
Press and hold the front panel TEST but-
ton for 5 s to initiate the following
sequence:
•
Audible Alarm Test
Audible alarm sounds and VFD
displays for 1 s.
Audible Test
•
LED Test
VFD displays for 9 s while other
modules perform LED tests.
LED Test
•
VFD Test
VFD scrolls the following message on
both rows
Test
ABCDEFGHIJKLMNOPQRSTUVWXYZ
Test
ABCDEFGHIJKLMNOPQRSTUVWXYZ
•
Checksum and IP Address Display
VFD displays checksum and device IP
address for 5 s.
Checksum: CCCC SSS
IP: XXX.XXX.XXX.XXX
C = 4 hex digit checksum
S = Ok if checksum is correct or
Bad if checksum is incorrect
X = IP Address
Signal Display
•
VFD displays each signal parameter for
2 s.
If the gas type for the signal is
the input gas type is displayed for 0.5 s.
Input XXX Gas Type
GGGGGGGGGGGGGGG
Input XXX Message
MMMMMMMMMMMMMMMMMMMM
Input XXX Location
LLLLLLLLLL
Input XXX
LED D
Unused