Monitoring Mode (Cont.)

Multi-Signal Module

Multi-Signal modules monitor up to five
dry-contact switches.
If the signal is normal (switch contacts
closed), green indicator for signal is
illuminated.
If an alarm occurs (switch contacts
open), audible alarm activates and red
indicator for signal flashes.
When front panel alarm MUTE
button is pressed, audible alarm is
silenced. Alarm indicator remains
illuminated, but does not flash.
Visual indicator will remain illuminated
as long as alarm condition remains.
If audible alarm repeat time has been
set, audible alarm will reactivate, and
visual indicator will again flash after
specified time.
Figure 96: Multiplexer Module
102
6-847718-00 Rev. B00
Signal 1
Signal 2
Signal 3
Signal 4
Signal 5