Multiplexer Module - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Multiplexer Module

Press and hold the front panel TEST but-
ton for 5 s to initiate the following
sequence:
Audible alarm sounds for 1 s.
Each green LED is displayed in
sequence for 0.5 s.
Each red LED is displayed in sequence
for 0.5 s.
Software revision is displayed as
follows:
-
Expressed as a three digit number
(e.g. 1.0.2).
-
Each digit of number is displayed
separately for 2 s and separated
from the next digit by a 1 s flashing
of all indicators.
-
Each number is displayed in a
binary code where each LED
indicator has a binary value
(Figure 109.).
-
Example: Software revision 1.0.2
would be displayed as follows:
First digit is determined by 2 s
illumination of top left LED (bit 0
= number 1).
All LEDs illuminate for 1 s.
Second digit is determined to be
0 because no LEDs are
illuminated for 2 s.
All LEDs illuminate for 1 s.
Third digit of is determined by 2
s illumination of 2nd red LED (Bit
1 = number 2).
After all three digits of software revision
have been displayed, all LEDs
illuminate for 1 s before displaying the
module's network address.
116
6-847718-00 Rev. B00
Network address is displayed for 2 s in
a binary code where each LED has a
binary value (Figure 109).
All LEDs illuminate for 1 s.
Module returns to monitoring mode.
Red LEDs
LED 1
Bit 0 (1)
LED 2
Bit 1 (2)
LED 3
Bit 2 (4)
LED 4
Bit 3 (8)
LED 5
Figure 109: Binary data represented by LEDs
Green LEDs
Bit 4 (16)
Bit 5 (32)
Bit 6 (64)
Bit 7 (128)