Digital Display Module

Press and hold the front panel TEST but-
ton for 5 s to initiate the following
sequence:
Audible alarm sounds for 1 s as T T E E S S T T is
displayed.
LED display segments and LOW,
NORMAL, and HIGH indicators illuminate
in sequence for 0.5 s.
Firmware revision (e.g. 2 2 . . 0 0 . . 0 0 ) is displayed
for 2 s.
A A D D D D R R is displayed for 1 s then module's
network address is displayed for 2 s.
Units of measure setting (e.g. P P S S I I ) is
displayed for 2 s.
L L O O is displayed for 1 s, LOW alarm LED is
illuminated, and current low alarm set point
is displayed for 2 s.
K K , , is displayed for 1 s, HIGH alarm LED is
illuminated, and current high alarm set
point is displayed for 2 s.
Module returns to monitoring mode.
117
6-847718-00 Rev. B00