Set-Up Procedures (Cont.)
Multi-Signal Module-
Setting Audible Alarm Repeat Time and
Display Intensity:
Used to set time interval for reactivation of
audible alarm after alarm MUTE
has been pressed.
1. Press and HOLD SETUP button.
Audible alarm will chirp when SETUP
button is first pushed, again after two
seconds and a third time after four
seconds. Release button after third
audible alarm chirp. Multi-signal module
front panel LED indicators will show
current audible alarm repeat time as
shown below:
Indicator Illuminated Repeat Time
None
Signal 1 - red
Signal 2 - red
Signal 3 - red
Signal 4 - red
Signal 1 - green
Signal 2 - green
Signal 3 - green
Signal 4 - green
Signal 5 - green
2. Using either
or
desired audible alarm repeat time.
NOTE:
To comply with the requirements of the City of
New York Department of Buildings; Materials
and Equipment Acceptance Division (M.E.A.),
set audible alarm repeat time to 30 minutes.
3 Press and release TEST button on front
of alarm panel. All LED indicators will
illuminate.
4. Select desired display brightness by
pressing either
push will increase or decrease
brightness of display. Audible alarm will
chirp one time with each button push
and two times when limit is reached.
Advanced (Cont.)
button
OFF
1 Min.
5 Min.
15 Min.
30 Min.
1 Hr.
2 Hr.
8 Hr.
12 Hr.
24 Hr.
buttons, select
or
buttons. Each
5. Press and release TEST button to save
changes. LEDs will sweep from top to
bottom to indicate end of set-up
sequence.
6. Multi-signal module will return to
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will default
to previous settings. No changes to settings will
be saved.
Set-up must be completed in order for changes
to be saved.