Operation; Start-Up And Checking - BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions

Medical electronic gas alarm
Table of Contents

Advertisement

Operation

Start-Up and Checking

1. Turn on electrical power to alarm
panel.
2. Make following observations:
POWER ON light illuminates on
front of alarm panel.
Digital display and multi-signal
modules warm-up for ten seconds.
During warm-up, no audible or
visual alarms activated. After
warm-up, audible and visual
alarms are initiated for any active
alarm.
3. Press MUTE
audible alarm.
4. Pressurize piping system (medical gas
and vacuum).
5. Make following observations:
Digital display modules display
actual pipeline pressures or
vacuum levels.
If digital display module pipeline
pressures and vacuum levels are
within pre-set limits, NORMAL
indicator is illuminated.
If digital display module pipeline
pressures and vacuum levels are
outside pre-set limits, status
indicators show HIGH or LOW.
If a fault code is displayed (- - F F 1 1 - - ,
- - F F 2 2 - - , - - F F 3 3 - - or - - F F 4 4 - - ) refer to
Troubleshooting Guide.
Multi-signal modules indicate
current status from dry-contact
switches. Unused signals can be
disabled. Refer to Set-Up
Procedure.
34
Part No. 6-847684-00 Rev. D01
button to silence
Status LED
Indicators
LED Display
Digital Display Module
Status LED
Indicators
Multi-Signal Module
NOTE:
If any module appears to be malfunctioning, refer
to Troubleshooting Guide.
SIGNAL #1
SIGNAL #2
SIGNAL #3
SIGNAL #4
SIGNAL #5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents