Option 014 Configurable Test Set - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A

Option 014 Configurable Test Set

The Option 014 analyzer allows you to measure devices with higher power and higher
dynamic range limits than the standard analyzer. The theory of operation is the same as
for the standard analyzer except that there are six front panel SMA jumpers.
As shown in
Figure 5-4 on page
reference signals between:
• the A22 switch splitter and the A25 test port 1 coupler. (For Option UNL, this is
between the A36 step attenuator and the A38 bias tee.)
• the A22 switch splitter and the A26 test port 2 coupler, (For Option UNL, this is
between the A37 step attenuator and the A39 bias tee.)
• the A25 test port 1 coupler and the A27 receiver A first converter
• the A26 test port 2 coupler and the A30 receiver B first converter
• the A23 detector (for E8363A and E8364A) or fixed attenuator (for E8362A) and the
A28 receiver R1 first converter
• the A24 detector (for E8363A and E8364A) or fixed attenuator (for E8362A) and the
A29 receiver R2 first converter
Normal Option 014 Configuration
The normal configuration for the Option 014 is shown in
configuration and inclusion of an external amplifier and accessories, you can calibrate the
analyzer and test devices at power levels up to +30 dBm. You can make measurements in
the forward, reverse or both directions and still achieve these high power levels. For more
information on higher power measurements, search for "Option 014" in the embedded help
index in the analyzer.
Service Guide E8364-90001
5-18, these jumpers allow access to the test port and
Theory of Operation
Signal Separation Group Operation
Figure 5-4 on page
5-18. With this
5- 17

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents