Removing And Replacing The A35 Receiver Motherboard - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the A35 Receiver Motherboard

Removing and Replacing the A35 Receiver Motherboard
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-12
1. Disconnect the power cord.
2. Remove the outer covers. Refer to
analyzer bottom side up and raise the receiver deck as shown. Refer to
Receiver Deck" on page 7-8
3. Remove all four receiver modules. Refer to
and A34 Receiver Modules" on page
4. Disconnect the two flexible cables (item
5. Disconnect the ribbon cable (item
6. Remove the receiver deck from the analyzer with the receiver motherboard attached.
Refer to
"Removing the Receiver Deck" on page
motherboard and receiver deck are replaced as one unit.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-52
for this procedure.
"Removing the Covers" on page
if necessary.
7-50.
) from the receiver motherboard.
Table 7-2 on page
PNA Series Microwave Network Analyzers
"Removing and Replacing the A31, A32, A33,
) from the receiver motherboard.
7-8, if necessary. The receiver
7-70.
E8362A, E8363A, E8364A
7-6. Position the
"Raising the
Service Guide E8364-90001