If Error Terms Seem Worse Than Expected - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
Table A-1
Error Term Data
Parameter
Directivity
Source Match
Standard
Option UNL, 014,
or UNL & 014
Load Match
Standard
Option UNL, 014,
or UNL & 014
b
Reflection Tracking
Transmission
b,c
Tracking
Crosstalk (Isolation)
Standard
Option UNL or 014
Option UNL & 014
a. Uncorrected system performance. The values in the table apply over an environmental temperature
of 23 °C ±3 °C, with less than 1 °C deviation from the calibration temperature.
b. Typical performance.
c. Transmission tracking performance is strongly dependent upon the cable used. These typical
specifications are set based on the use of Agilent through cable, part number 85133-60016.
d. The value is −85 dB for frequencies below 1 GHz.

If Error Terms Seem Worse than Expected

To verify that the system still conforms to specifications, perform a system verification.
Refer to
Chapter 3, "Tests and
Service Guide E8364-90001
a
45 MHz to
2 GHz to
2 GHz
10 GHz
>24 dB
>22 dB
>23 dB
>16 dB
>18 dB
>14 dB
>23 dB
>14 dB
>17 dB
>13 dB
±1.5 dB
±1.5 dB
±1.5 dB
±2.0 dB
d
−100 dB
−110 dB
−109 dB
−100 dB
d
−108 dB
d
−98 dB
Adjustments".
Frequency Range
10 GHz to
20 GHz to
20 GHz
40 GHz
>16 dB
>16 dB
>14 dB
>10 dB
>12 dB
>8 dB
>10 dB
>9 dB
>10 dB
>9 dB
±1.5 dB
±1.5 dB
±2.5 dB
±3.5 dB
−110 dB
−108 dB
−109 dB
−106 dB
−107 dB
−104 dB
Error Terms
Error Term Data
40 GHz to
45 GHz to
45 GHz
50 GHz
>15 dB
>13 dB
>9 dB
>5.5 dB
>7 dB
>4 dB
>9 dB
>7 dB
>8 dB
>6 dB
±2.0 dB
±2.0 dB
±4.0 dB
±4.5 dB
−105 dB
−100 dB
−103 dB
−98 dB
−100 dB
−95 dB
A- 9