PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
8. The log magnitude ratioed measurement using the receiver under test is read: (P
9. The source output level is adjusted to decrease the output by 15 dB.
10. The log magnitude ratioed measurement using the receiver under test is read: (P
11. Attenuation of 20 dB is introduced between the source and receiver. This can be done
with an internal step attenuator, an external 10 dB pad (if a short is being used to
reflect the signal back into the port), or an external 20 dB pad (if the signal source is
the compression test set).
12. The log magnitude ratioed measurement using the receiver under test is read: (P
13. The source output level is adjusted to produce a reading of Pref (within ±0.03 dBm) for
the reference channel on the source port.
14. The log magnitude ratioed measurement using the receiver under test is read: (P
15. The compression in dB = (P
If the Analyzer Fails this Test:
• Repeat this test. There are no adjustments that can be made.
• Since all four receivers are identical, the suspect failed receiver can be swapped with
one of the other receivers to verify that it is faulty. Swap the suspect receiver and repeat
this test. Refer to
Modules" on page
• If the faulty channel "follows" the receiver, replace the receiver. If the same channel
(now with a different receiver) is still faulty, suspect the associated cabling.

Noise Floor Test

Function of the Test: To measure the absolute power level of the noise floor for the
analyzer's receivers.
Specification Tested: Test Port Input–Test Port Noise Floor
Equipment Used: A power meter, power sensor, a calibration kit, and a test cable.
Description of the Test:
1. The analyzer is set to various CW frequencies across its frequency range at an IF
bandwidth of 1 kHz and 801 points per sweep.
2. A test cable is connected to the driving port for the measurement and a power sensor is
connected to the other end of the cable.
3. The power level at the end of the cable is set to −5.00 dBm.
4. The power sensor is disconnected and the cable is connected to the port to be tested.
5. The absolute power level in dBm (log magnitude) is read: (P
6. The test cable is removed and loads are connected to both ports.
7. The analyzer's trace is set to represent the absolute power level (linear magnitude) for
the receiver under test and a sweep is taken.
Service Guide E8364-90001
−P
−P
)−(P
a
b
d
"Removing and Replacing the A31, A32, A33, and A34 Receiver
7-50.
Performance Tests (Agilent N2721A Software Package)
).
c
log
Tests and Adjustments
).
a
).
b
).
c
).
d
).
3- 33