Agilent Technologies E8362A Service Manual page 157

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
Theory of Operation
E8362A, E8363A, E8364A
Signal Separation Group Operation
High Dynamic Range Configuration
With a few jumper changes, you can configure the Option 014 for higher dynamic range
measurements. By swapping the front panel jumpers for one port, signal flow through the
corresponding coupler is reversed, increasing the test signal sensitivity by 15 dB.
In the forward direction, as shown in
Figure
5-5, the signal flow through the test port
coupler (A26) is reversed by arranging the front panel jumpers such that RCVR B IN
connects to CPLR THRU and CPLR ARM connects to SOURCE OUT.
In the reverse direction, not shown, the signal flow through the test port coupler (A25) is
reversed by arranging the front panel jumpers such that RCVR A IN connects to
CPLR THRU and CPLR ARM connects to SOURCE OUT.
For more information on high dynamic range measurements and configurations, search for
"Option 014" in the embedded help index in the analyzer.
Figure 5-5 High Dynamic Range Configuration in the Forward Direction
Service Guide E8364-90001
5- 19

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents