PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
c. Adjust the horizontal width of the Windows Explorer window to approximately
one-half the width of the display and position it to either the left or right side of the
display.
d. Repeat this procedure to create another copy of the Windows Explorer window and
position it to the opposite side of the display so that the two copies are displayed
side-by-side.
3. Set one copy of Windows Explorer to the A: drive and the other to the C: drive:
a. In one copy of Windows Explorer, double-click on My Computer, and then
double-click on 3 1/2 Floppy (A:). A listing of the files on the floppy disk will appear
in the right-hand side of the window.
b. In the other copy of Windows Explorer, double-click on My Computer, and then
double-click on Hard Disk (C:).
c. Navigate to C:\Program Files\Agilent\Network Analyzer by double-clicking on
each of the names in this path name, in turn. A listing of the files in this directory on
the hard disk will appear in the right-hand side of the window.
4. Copy each of the backup files from the floppy disk to the hard disk drive:
a. Click and drag each of the files from the A: drive to the C:\Program
Files\Agilent \Network Analyzer directory by clicking and holding on the file
name in the A: directory and, while holding the mouse button down, drag the file
name to the Network Analyzer directory on the C: drive. If you need help with this
"dragging" process, refer to Windows Help. (Click START, HELP. Click on the Index
tab and type dragging in the dialog box.)
5. After all files have been copied, remove the floppy disk from the analyzer's disk drive
and exit both copies of Windows Explorer.
Set Up the Agilent Administrator Password
If the hard disk has been replaced, it is necessary to set up the Agilent administrator
password for the analyzer. This requires obtaining a password from Agilent:
Obtaining a Password
• If this procedure is being performed by Agilent personnel, go to the following Internet
site and follow the instructions under "Agilent Administrator Password."
http://mktwww.soco.agilent.com/field/service/network/pna/support.htm
• If this procedure is being performed by non-Agilent personnel, contact Agilent for a
password before beginning this procedure. Refer to
Entering Your Agilent Administrator Password
The network analyzer must be powered up and operating for this procedure.
1. Exit the Network Analyzer program, if it is running.
2. On the Windows Desktop, right-click on My Computer, and then click Manage.
3. In the Computer Management window, navigate to System Tools, Local Users and
Groups, Users.
Service Guide E8364-90001
Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
"Contacting Agilent" on page
2-9.