PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
• all user accounts and passwords
• all added programs
• all saved instrument states
• all user configurations and Windows settings
The above listed information can be saved by copying it to a floppy disk or to a folder on the
D: drive; the D: drive will not be affected by the restoration process. In addition, factory
instrument calibration will not be affected if no changes have been made since the original
factory calibration. If changes have been made through the service routines, copy all files
that are prefixed with mxcalfile_ in the C:/Program Files/Agilent/Network Analyzer directory
to the D:/Calfiles directory. Permanent instrument options should not be lost.
Service Guide E8364-90001
Operating System Recovery
Overview
D- 3