Removing And Replacing The A21 Soma 50 (E8363A And E8364A Only) - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the A21 SOMA 50 (E8363A and E8364A Only)

Removing and Replacing the A21 SOMA 50 (E8363A and
E8364A Only)
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-13
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
analyzer bottom side up and raise the receiver deck as shown. Refer to
Removing the Receiver Deck" on page 7-8
3. With a 5/16-inch torque wrench, disconnect the two semirigid cables (item
A21 SOMA 50.
4. Disconnect the flexible cable (item
5. Disconnect the ribbon cable (item
6. With a T-10 TORX driver, remove the four mounting screws (item
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing
of the cables is necessary to remove the assembly. Do not over-bend the
semirigid cables.
7. Remove the A21 SOMA 50 from the analyzer. Observe the
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-38
for this procedure.
"Removing the Covers" on page
) from the A21 SOMA 50.
) at the A21 SOMA 50.
Table 7-2 on page
PNA Series Microwave Network Analyzers
if necessary.
CAUTION
7-70.
Service Guide E8364-90001
E8362A, E8363A, E8364A
7-6. Position the
"Raising and
) from the
).
above.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents