Agilent Technologies E8362A Service Manual page 39

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
Accessories Used in the Operator's Check
Equipment Type
Female short, 3.5 mm
Female short, 2.4 mm
Performing the Operator's Check
1. From the System menu, point to Service, and then click Operator's Check.
2. In the Operator's Check dialog box (refer to
select either Automatic, to run through the test without stopping, or Prompted, to
pause at each step in the process.
3. To check Port 1, click Start--Port 1 or to check Port 2, click Start--Port 2.
4. The test will prompt you to connect the short.
5. The result of the operator's check will be shown as a PASS or FAIL next to the test.
Figure 3-3
Operator's Check Dialog Box
If the Operator's Check Fails
1. Clean the test ports, shorts, and adapters. Torque to specification. Repeat the check.
2. If the check still fails, suspect a faulty component:
• A repeated failure in the switch repeatability for both the Port 1 and Port 2 checks
indicates a faulty A22 switch splitter. Refer to
Switch Splitter" on page
Service Guide E8364-90001
Models Used With
E8362A
E8363A, E8364A
7-40.
Part Number
85052-60007 (from the 85052 calibration kit)
85056-60021 (from the 85056 calibration kit)
Figure
3-3), click Configure, and then
"Removing and Replacing the A22
Tests and Adjustments
Preliminary Checks
3- 11