Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
Replacement Procedure
WARNING
When connecting the hard disk (item
interface board, take special care to align the connector pins.
Applying power to a misaligned connection can permanently damage
the hard disk drive.
When installing the hard disk, first torque all four mounting screws (item
IMPORTANT
in
Figure
7 in-lb with a T-10 TORX driver.
Installing the New A41 Hard Disk Drive
Refer to
Figure 7-30 on page
1. With the disk drive bracket right-side up (it is shown upside down in
page
7-63), insert four bushings (item
2. Connect the new hard disk drive (item
firmly into the connector on the interface board. Refer to the
3. While holding the disk drive in place, turn the bracket over and insert the other four
bushings (item
4. Install the four screws (item
attach the hard disk drive to the bracket.
With a T-10 TORX driver, torque the four screws, in a Z-pattern, to 4 in-lb, and then
torque them again, in the same pattern, to 7 in-lb. Refer to the
Reinstalling the HDDA
Refer to
Figure 7-29 on page
1. Slide the hard disk drive assembly into the slot in the rear panel and push it firmly into
place to engage the connector.
2. With a T-10 TORX driver, tighten the four HDDA mounting screws (item
Install Backup Files onto the New Hard Disk Drive
The files that were saved onto a floppy disk in Step 1 must now be installed onto the new
hard disk drive for the analyzer to operate properly.
The network analyzer must be powered up and operating for this procedure.
1. Insert the floppy disk containing the backup files into the floppy disk drive.
2. Open two copies of Windows Explorer:
a. Click on START. (It may be necessary to exit the Network Analyzer application.)
b. Point to Programs, Accessories, and then click on Windows Explorer.
7-64
7-30) to 4 in-lb with a T-10 TORX driver. Then secure each screw to
7-63.
), into the other side of the same four holes specified in step 1.
), finger tight, through the eight bushings (item
7-62.
PNA Series Microwave Network Analyzers
in
Figure
) into the four holes in the bracket.