Agilent Technologies E8363A Installation Note

Configurable test set upgrade kit pna series microwave network analyzers
Hide thumbs Also See for E8363A:

Advertisement

Quick Links

Installation Note
Configurable Test Set Upgrade Kit
For E8363A and E8364A PNA Series Microwave Network Analyzers
Network Analyzer
Upgrade Kit
Model Number
Part Number
E8363A, E8364A
E8364-60101
Agilent Part Number: E8364-90012
Printed in USA January 2005
Supersedes print date: April 2004
© Agilent Technologies, Inc. 2002, 2004, 2005

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies E8363A

  • Page 1 Installation Note Configurable Test Set Upgrade Kit For E8363A and E8364A PNA Series Microwave Network Analyzers Network Analyzer Upgrade Kit Model Number Part Number E8363A, E8364A E8364-60101 Agilent Part Number: E8364-90012 Printed in USA January 2005 Supersedes print date: April 2004...
  • Page 2: Warranty Statement

    “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3 Getting Assistance from Agilent By internet, phone, or fax, get assistance with all your test and measurement needs. This information supersedes all prior HP contact information. Online assistance: www.agilent.com/find/assist Americas Brazil Canada Mexico United States (tel) 1 800 254 2440 (tel) (+55) 11 3351 7012 (tel) +1 877 894 4414 (tel) 800 829 4444...
  • Page 4 About Installing the Upgrade Kit Products affected......E8363A and E8364A; all options Installation to be performed by ....Agilent service center or personnel qualified by Agilent Estimated installation time .
  • Page 5 Items Included in the Upgrade Kit Table 1 lists the parts included in this upgrade kit, Agilent part number E8364-60101. Check the contents of your kit against this list. If any item is missing or damaged, contact Agilent Technologies. Refer to “Getting Assistance from Agilent”...
  • Page 6: Electrostatic Discharge Protection

    Installation Procedure for the Upgrade Kit The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation. This installation requires the removal of the analyzer’s protective outer WARNING covers.
  • Page 7: Overview Of The Installation Procedure

    Overview of the Installation Procedure Step 1. Remove the Outer Cover Step 2. Remove the Front Panel Assembly Step 3. Raise the Receiver Deck Step 4. Remove the Existing Cables Step 5. Install the Option 014 Cables Step 6. Lower and Fasten the Receiver Deck Step 7.
  • Page 8 Step 1. Remove the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. Refer to Figure 2 for this procedure.
  • Page 9: Step 2. Remove The Front Panel Assembly

    Step 2. Remove the Front Panel Assembly Refer to Figure 3 for this procedure. 1. With a T-10 TORX driver, remove the eight screws (item ) from the sides of the frame. Before removing the front panel from the analyzer, lift and support the front of CAUTION the analyzer chassis.
  • Page 10 Step 3. Raise the Receiver Deck Refer to Figure 4 for this procedure. 1. Place the analyzer bottom-side up on a flat surface. 2. With a T-10 TORX driver, remove the four screws (item ), securing the receiver deck. 3. Pull the latch pin (item ) towards the center of the analyzer to release the receiver deck.
  • Page 11 Step 4. Remove the Existing Cables Analyzers without Option UNL Refer to Figure 5 for the following procedure. If you are installing Option 014 on an analyzer that does not have Option UNL installed, remove the following cables in the order listed: •...
  • Page 12 Analyzers with Option UNL Refer to Figure 6 for the following procedure. If you are installing Option 014 on an analyzer that has Option UNL installed, remove the following cables in the order listed: • W8 E8364-20080 A24 detector to A29 channel R2 mixer •...
  • Page 13 Step 5. Install the Option 014 Cables Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections in this step CAUTION of the procedure. Analyzers without Option UNL Refer to Figure 7 for the following procedure. The new parts referenced in this procedure are listed in Table 1 on page If you are installing Option 014 on an analyzer that does not have Option UNL installed,...
  • Page 14 Analyzers with Option UNL Refer to Figure 8 for the following procedure. The new parts referenced in this procedure are listed in Table 1 on page If you are installing Option 014 on an analyzer that has Option UNL installed, install the following cables in the order listed: •...
  • Page 15 Step 6. Lower and Fasten the Receiver Deck Refer to Figure 9 for this procedure. 1. Pull the latch pin (item ) toward the center of the analyzer to release the receiver deck. 2. Lift the receiver deck to partially lower it, then release the latch pin (item ).
  • Page 16 Step 7. Replace the Lower Front Panel Overlay The new parts referenced in this procedure are listed in Table 1 on page NOTE Refer to Figure 10 for this procedure. 1. From the back side of the front panel, use a blunt object in one of the cutouts in the lower frame to push the overlay (item ) and separate it from the front panel.
  • Page 17 Step 8. Reinstall the Front Panel Assembly and Install the Front Panel Jumpers Before installing the front panel assembly onto the analyzer, lift and support the CAUTION front of the analyzer chassis. Refer to Figure 11 for this procedure. 1. Tighten all 12 of the front-panel feed-through connectors using a 5/16-in torque wrench set to 21-in lbs.
  • Page 18 Step 9. Reinstall the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. Refer to Figure 12 for this procedure.
  • Page 19 2. Verify that “014” is listed after “Options:” in the display. 3. Click OK. If Option 014 has not been enabled, perform “Enable Option 014” again. If the NOTE option is still not enabled, contact Agilent Technologies. Refer to “Getting Assistance from Agilent” on page Installation Note E8364-90012...
  • Page 20 Step 11. Perform Post-Upgrade Adjustments and Calibration The following adjustments must be made due to the change in the full frequency range of the analyzer. • source calibration • receiver calibration These adjustments are described in the PNA service guide and in the PNA on-line HELP. A list of equipment required to perform these adjustments can be found at “Equipment Required for Post-Upgrade Adjustments”...