Receiver Calibration Adjustment - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

Tests and Adjustments
Adjustments

Receiver Calibration Adjustment

The receiver calibration is used to adjust the network analyzer receivers for a flat response
across its full frequency range:
1. A power meter/sensor is connected to Port 1, as shown in
reference for flatness.
2. A cable is inserted between the power sensor and the test port, as shown in
to establish a reference for the cable.
3. The same cable is connected between the two test ports, as shown in
signal from Port 1 is used to adjust the "B" receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the "A" receiver at
Port 1.
Data obtained during this adjustment are stored in the mxcalfile_pxx files on the hard disk
drive (where xx is a 1 or 2 digit number). The data are used in subsequent measurements.
If the hard disk drive is replaced, these mxcalfile_pxx files will be lost. Therefore, they
should be backed up (saved on a floppy disk) so that they can be restored.
These files can be recreated by performing another receiver calibration adjustment.
Equipment Used for the Receiver Calibration Adjustment
Equipment Type
Power meter
Power sensor, 3.5 mm
Adapter, 3.5 mm (f) to 3.5 mm (f)
RF Cable, 3.5 mm (f) to 3.5 mm (f)
Power sensor, 2.4 mm
Adapter, 2.4 mm (f) to 2.4 mm (f)
RF Cable, 2.4 mm (f) to 2.4 mm (f)
3-44
Models
Model or
Used With
Part Number
All
E4418B/E4419B
E8362A
E4413A
E8362A
83059B
E8362A
85131C
E8363A, E8364A
8487A
E8363A, E8364A
11900B
E8363A, E8364A
85133C
PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
Figure
3-19, to establish a
Figure
Figure
3-21, and a
Alternate Model
Part Number
E4418A/E4419A
8485A
85052-60012
85131E
None
85056-60007
85133E
Service Guide E8364-90001
3-20,

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents