Option Enable Utility
Installing or Changing a Serial Number
Installing or Changing a Serial Number
It is necessary to reinstall the instrument serial number if the A16 test set motherboard is
replaced.
The instrument serial number is printed on a tag located in the lower rear corner of the
right-hand side of the outer cover.
IMPORTANT
Use extreme care when entering the serial number, as only one attempt is
allowed.
1. To change a serial number, select Change Serial from the Option Enable menu bar (see
Figure
B-1). The current serial number is displayed. If no serial number has previously
been entered, the word "NONE" will be displayed.
2. Type the new serial number into the space provided, click Serial.
NOTE
To change an incorrect serial number, a clear-code password is required.
Contact Agilent to obtain the clear-code. Refer to
page
2-9.
B-6
PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
"Contacting Agilent" on
Service Guide E8364-90001