Agilent Technologies E8362A Service Manual page 53

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
System Verification Procedure
1. Connect a cable to Port 1 on the analyzer as shown in
test outputs, connect a printer to the analyzer. For the printer, ensure that the correct
driver is loaded and the printer is defined as the default printer. Refer to the embedded
help in the analyzer for printer setup. Let the analyzer warm up for at least 90 minutes.
Figure 3-11 System Verification Test Setup
2. Insert the verification kit disk into the analyzer disk drive.
3. On the System menu, point to Service, and then click System Verification. The System
Verification dialog box is displayed; refer to
Figure 3-12 System Verification Dialog Box
Service Guide E8364-90001
Figure
3-11. If you desire printed
Figure
3-12.
Tests and Adjustments
System Verification
3- 25

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363aE8364a

Table of Contents