Tools Required - Agilent Technologies E8362A Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362A, E8363A, E8364A
b. Click and drag each of the files on the C: drive from the C:\Program
Files\Agilent\Network Analyzer directory by clicking and holding on the file
name in the directory and, while holding the mouse button down, drag the file name
to the A: drive. If you need help with this "dragging" process, refer to Windows Help.
(Click START, HELP. Click on the Index tab and type dragging in the dialog box.)
5. After all files have been copied, remove the floppy disk from the analyzer's disk drive
and exit both copies of Windows Explorer.

Tools Required

• T-10 TORX driver (set to 4 in-lb; for hard disk drive replacement)
• T-10 TORX driver (set to 7 in-lb; for hard disk drive replacement)
• T-10 TORX driver (set to 9 in-lb; for all other T-10 applications)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Overview of Procedures
There are two procedures related to removing and replacing the hard disk drive assembly
(HDDA):
Removing the HDDA
Disconnect and remove the HDDA from the analyzer so that the following assemblies may
be serviced.
• A14 system motherboard
• A16 test set motherboard
• A41 hard disk drive
Disassembling the HDDA
This procedure outlines how to disassemble the HDDA, once it has been removed from the
analyzer, to replace the A41 hard disk drive.
Service Guide E8364-90001
Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
7- 61