Troubleshooting
Measurement System Troubleshooting

Checking the Receiver Group

Equipment Used for These Tests
Equipment Type
Spectrum analyzer
1. Must be capable of measuring signals at 8.333 MHz and 41.667 MHz.
Getting Ready to Test
Before checking the assemblies, you must open the analyzer.
CAUTION
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
1. Turn off the analyzer power.
2. Unplug the power to the analyzer and disconnect all rear panel connections.
3. Remove the covers from the analyzer. Refer to
4. With the covers off, plug in the analyzer and turn on the power.
Checking the A27 through A30 First Converter Outputs
Perform this procedure if you observe a problem in all bands on a single channel.
1. Refer to the block diagram at the end of this chapter and to
Standard" on page
converters, A27, A28, A29, and A30, respectively.
2. Using a 5/16-inch torque wrench, disconnect the semirigid cable associated with the
faulty channel, from the first converter.
3. Connect the spectrum analyzer to the connector to which this cable was connected.
4. The measured signal on the spectrum analyzer should be at 8.333 MHz.
5. If the signal is not present, replace the faulty first converter. Refer to
Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias Board"
on page
7-48.
6. If the signal is present, reconnect the cable and continue testing at
through A34 Receiver
Checking the A31 through A34 Receiver Outputs
Perform this procedure if a problem appears to be in one of the receiver channels.
1. Refer to the block diagram at the end of this chapter and to
Standard" on page
receiver IF output.
2. Disconnect the flexible cable at the suspect receiver.
4-50
Models
Used With
All
6-26. Locate the cables W21, W22, W23, and W24 at the first
Outputs".
6-26. Locate the flexible cables W25, W26, W27 and W28 at each
PNA Series Microwave Network Analyzers
Model or
Part Number
8565E
"Removing the Covers" on page
"Bottom RF Cables,
"Bottom RF Cables,
Service Guide E8364-90001
E8362A, E8363A, E8364A
Alternate Model
or Part Number
1
856xE
7-6.
"Removing and
"Checking the A31