Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive
4. Right-click on Agilent, and then click on Set Password.
a. Type the password obtained from Agilent as the Password. Refer to
Password" on page 7-65
b. Type the password again in the Confirm password block.
c. Click OK.
5. Right-click on Agilent, and then click on Properties.
a. Unselect User must change password at next logon (if not already unselected).
b. Unselect User cannot change password (if not already unselected).
c. Select Password never expires (if not already selected).
d. Click OK.
6. Close the Computer Management window.
Post-Repair Procedures
Perform the post-repair adjustments, verifications, and performance tests that pertain to
this removal procedure. Refer to
7-66
for information if you have not yet obtained this password.
Table on page
7-70.